[{"data":1,"prerenderedAt":-1},["ShallowReactive",2],{"doc-seo-398082-105":3,"detail-sidebar-cat-0-en-105":81,"doc-detail-398082-en":131},{"code":4,"msg":5,"data":6},0,"ok",{"site_id":7,"language":8,"slug":9,"title":10,"keywords":11,"description":12,"schema_data":13,"social_meta":74,"head_meta":76,"extra_data":78,"updated_unix":80},105,"en","bgia-report-test-report-no-2008214692120","BGIA-Report - Test Report - No. 200821469/2120","","Formal test report by BGIA-PRÜFZERT dated 10 April 2008 evaluating the automated filter testing device TSI 8130. The development test assesses filter penetration measurements using aerosol generators and compares results after adjusting the aerosol particle size distribution to the BGIA-CEN standard. Adjustments reduced the conversion ratio for both paraffin oil and NaCl aerosols. A final suitability statement for type approval and full comparability is pending due to limited comparative values and relatively high standard deviations.",{"@graph":14,"@context":73},[15,34,56],{"@type":16,"itemListElement":17},"BreadcrumbList",[18,23,27,31],{"item":19,"name":20,"@type":21,"position":22},"https://docshare.wps.com","Home","ListItem",1,{"item":24,"name":25,"@type":21,"position":26},"https://docshare.wps.com/document/","Document",2,{"item":28,"name":29,"@type":21,"position":30},"https://docshare.wps.com/document/research-report/","Research & Report",3,{"item":32,"name":10,"@type":21,"position":33},"https://docshare.wps.com/document/bgia-report-test-report-no-2008214692120/398082/",4,{"url":32,"name":10,"@type":35,"image":36,"author":41,"headline":10,"publisher":44,"fileFormat":47,"inLanguage":8,"description":12,"dateModified":48,"datePublished":49,"encodingFormat":47,"isAccessibleForFree":50,"interactionStatistic":51},"DigitalDocument",{"url":37,"@type":38,"width":39,"height":40},"https://docshare.wps.com/thumbnails/bgia-report-test-report-no-2008214692120/398082.png","ImageObject",300,407,{"name":42,"@type":43},"Franzy","Person",{"url":19,"name":45,"@type":46},"DocShare","Organization","application/pdf","2026-11-27","2026-09-27",true,{"@type":52,"interactionType":53,"userInteractionCount":55},"InteractionCounter",{"@type":54},"ViewAction",11,{"@type":57,"mainEntity":58},"FAQPage",[59,65,69],{"name":60,"@type":61,"acceptedAnswer":62},"What is the purpose of this BGIA test report?","Question",{"text":63,"@type":64},"It evaluates the suitability of the automated filter testing device TSI 8130 by measuring particle filter penetration and comparing it to BGIA/CEN test equipment results.","Answer",{"name":66,"@type":61,"acceptedAnswer":67},"How were test results made comparable?",{"text":68,"@type":64},"The aerosol particle size distribution of the TSI 8130 aerosol generators was adjusted to match the BGIA-CEN aerosol particle size distribution, making the measured filter penetrations comparable.",{"name":70,"@type":61,"acceptedAnswer":71},"Why is a final statement about type approval suitability not yet possible?",{"text":72,"@type":64},"Because the number of comparative values is currently low and the achieved standard deviations of the conversion factors are relatively high, including variance from the test rig and the tested filter samples.","https://schema.org",{"og:url":32,"og:type":75,"og:title":10,"og:site_name":45,"og:description":12},"article",{"robots":77,"canonical":32},"index,follow",{"doc_id":79,"site_id":7},398082,1790578924,{"code":4,"msg":82,"data":83},"success",[84,88,92,96,101,106,111,115,120,123,127],{"id":22,"doc_module":4,"doc_module_name":25,"category_name":85,"show_sort_weight":86,"slug":87},"Story & Novel",90,"story-novel",{"id":26,"doc_module":4,"doc_module_name":25,"category_name":89,"show_sort_weight":90,"slug":91},"Literature",80,"literature",{"id":33,"doc_module":4,"doc_module_name":25,"category_name":93,"show_sort_weight":94,"slug":95},"Exam",70,"exam",{"id":97,"doc_module":4,"doc_module_name":25,"category_name":98,"show_sort_weight":99,"slug":100},5,"Comic",60,"comic",{"id":102,"doc_module":4,"doc_module_name":25,"category_name":103,"show_sort_weight":104,"slug":105},6,"Technology",50,"technology",{"id":107,"doc_module":4,"doc_module_name":25,"category_name":108,"show_sort_weight":109,"slug":110},7,"Healthcare",40,"healthcare",{"id":112,"doc_module":4,"doc_module_name":25,"category_name":29,"show_sort_weight":113,"slug":114},8,30,"research-report",{"id":116,"doc_module":4,"doc_module_name":25,"category_name":117,"show_sort_weight":118,"slug":119},9,"Religion & Spirituality",20,"religion-spirituality",{"id":118,"doc_module":4,"doc_module_name":25,"category_name":121,"show_sort_weight":118,"slug":122},"World Cup","world-cup",{"id":124,"doc_module":4,"doc_module_name":25,"category_name":125,"show_sort_weight":124,"slug":126},10,"Lifestyle","lifestyle",{"id":128,"doc_module":4,"doc_module_name":25,"category_name":129,"show_sort_weight":97,"slug":130},19,"General","general",{"code":4,"msg":82,"data":132},{"doc_id":79,"user_id":133,"nickname":42,"user_avatar":134,"doc_module":4,"category_id":112,"category_name":29,"doc_title":10,"doc_description":12,"doc_content":135,"file_id":136,"file_url":137,"file_type":138,"file_size":139,"view_count":55,"is_deleted":4,"is_public":22,"is_downloadable":22,"audit_status":22,"page_count":112,"language":140,"language_code":8,"site_id":7,"html_lang":8,"table_of_contents":141,"faqs":142,"seo_title":143,"seo_description":12,"update_tm":144,"read_time":118,"preview_page_count":33,"preview_urls":145},2336478945519,"https://ap-avatar.wpscdn.com/davatar_085a072bc5b1113ac321206ff7593b45","Prüf-und Zertifizierungsstelle  \nIm BG-PRÜFZERT  \nDate: 10th of April 2008 Tob/The  \nTEST REPORT  \n- This translation into English was made by TSI  \nNo.: 200821469/2120  \n1 Customer TSI GmbH  \nNeuköllner Strasse 4  \n52068 Aachen Germany  \n2 Test specimen Automated Filter Tester  \n2.1 Manufacturer TSI  \n2.2 Type, designation Testing device for the determination of particle penetration through filters / TSI 8130  \nMarking TSI 8130  \n2.3 Intended use N.A.  \n2.4 Date of fabrication N.A.  \n2.5 Further details N.A.  \nD-53754 Sankt Augustin  \nTel.: (02241)231-02 Fax: (02241)231-2234 [email:](email: bgia@dguv.de)[ ](email: bgia@dguv.de)[bgia@dguv.de](email: bgia@dguv.de) b_de 09.07  \n3 Testing  \n3.1 Type of test Development test  \n3.2 Date of Testing November 2007  \n3.3 Test method, requirements according to DIN EN 143:2007  \n4 Assessment, suitability  \nThe comparative tests have shown that-after adjustment of the aerosol particle size distribution of the aerosol generators of the TSI 8130 to the BGIA-CEN aerosol particle size distribution – the filter penetration measured by the TSI 8130 approximates the penetration measured with the CEN-TC79 test equipment. The test results thus become comparable.  \nDue to the changes made at the TSI 8130, the conversion ratio TSI 8130/BGIA was reduced significantly from greater than 1.8 to 1.02 for paraffin oil aerosol and from greater than 1.6 to 1.15 for NaCl aerosol, respectively. This approximation is sufficient for production quality control testing.  \nA statistically valid final statement about the suitability of the TSI 8130 testing device for type approval tests of respiratory protection devices at test laboratories and the comparability of filter penetrations as determined by the TSI 8130 testing device with such penetrations determined according to CEN-TC97 is not yet possible due to the presently low number of comparative values and hereby achieved standard deviations of conversation factors of greater than 0. 1, which are relatively high. The standard deviation of the conversation factors includes the variance both due to the test rig as well as the tested filter samples.  \nA final assessment by BGIA can be made after confirming the present results based on further testing of a larger number of suitable particle filters with an accordingly modified TSI 8130 testing device.  \n5 Validity of Test Report  \nThe test results apply to the tested objects only.  \nLimitation of validity or use of this Test Report: -  \nD-53754 Sankt Augustin Tel.: (02241)231-02 Fax: (02241)231-2234 [email:](email: bgia@dguv.de)[ ](email: bgia@dguv.de)[bgia@dguv.de](email: bgia@dguv.de) b_de 09.07  \n6 General remarks  \nThe present Test Report consists of 8 pages  \nPages 1 to 3 indicate the overall test result. The complete Test Report also includes the test protocol containing all pertinent details.  \nThe present Test Report does n o t warrant the use of the GS-label, BG-label or CE-mark.  \nIn all other respects the Rules of Procedure for Testing and Certification carried out by the Test and Certification Bodies in BG-PRÜFZERT shall apply in conjunction with the General Business Conditions of the Deutsche Gesetzliche Unfallversicherung e.V.  \nFor the assessment:  \nDipl.Ing. Hans-Ulrich Tobys Certification officer  \nFor the testing:  \nClaudia Lietz  \nHead of Test laboratory  \nD-53754 Sankt Augustin  \nTel.: (02241)231-02  \nFax: (02241)231-2234 [email:](email: bgia@dguv.de)[ ](email: bgia@dguv.de)[bgia@dguv.de](email: bgia@dguv.de)  \nb_de  \n09.07  \nTest protocol  \n1 Test requirements: according to DIN EN 143:2007  \n2 Type of test: development test  \n3 Customer: TSI GmbH  \n4 Test specimen  \n4.1 Type: Testing device for the determination of particle filter penetration  \n4.2 Designation: TSI 8130  \n4.3 Marking: TSI 8130  \n5 Reason for Testing  \nThe TSI testing device for the determination of particle filter penetration, TSI 8130, is designed for the determination of particle filter penetration according to the American S","cbCaibo0jqcRMFyy","https://ap.wps.com/l/cbCaibo0jqcRMFyy","pdf",155761,"English","# 1 Customer\n## 2 Test specimen\n## 2.1 Manufacturer\n## 2.2 Type, designation\n## 2.3 Intended use\n## 3 Testing\n# 4 Assessment, suitability\n# 5 Validity of Test Report\n# 6 General remarks\n# Test protocol\n## 1 Test requirements\n## 2 Type of test\n## 3 Customer\n## 4 Test specimen\n## 5 Reason for Testing\n## 6 Operating parameters\n## 7 Particle filters for comparisons","[{\"question\":\"What is the purpose of this BGIA test report?\",\"answer\":\"It evaluates the suitability of the automated filter testing device TSI 8130 by measuring particle filter penetration and comparing it to BGIA/CEN test equipment results.\"},{\"question\":\"How were test results made comparable?\",\"answer\":\"The aerosol particle size distribution of the TSI 8130 aerosol generators was adjusted to match the BGIA-CEN aerosol particle size distribution, making the measured filter penetrations comparable.\"},{\"question\":\"Why is a final statement about type approval suitability not yet possible?\",\"answer\":\"Because the number of comparative values is currently low and the achieved standard deviations of the conversion factors are relatively high, including variance from the test rig and the tested filter samples.\"}]","BGIA-Report - Test Report - No. 200821469/2120 | PDF",1790470630,[146,150,154,158],{"page_no":22,"url":147,"width":148,"height":149},"https://res-academy.cache.wpscdn.com/upload/kdocshare/2026-11-16/425201d182a1563f1b9b6c6d3dd78275.webp",983,1272,{"page_no":26,"url":151,"width":152,"height":153},"https://res-academy.cache.wpscdn.com/upload/kdocshare/2026-11-16/b43707554adb340003922a377e1e7875.webp",982,1270,{"page_no":30,"url":155,"width":156,"height":157},"https://res-academy.cache.wpscdn.com/upload/kdocshare/2026-11-16/0b217361b4ce62a1874d18e74240d7df.webp",981,1269,{"page_no":33,"url":159,"width":160,"height":161},"https://res-academy.cache.wpscdn.com/upload/kdocshare/2026-11-16/e6d97ea3cc6b3c2c4ca5e57e63080526.webp",980,1268]